Endosafe
®
-
Portable Test System User’s Guide
Copyright 2003
Page A2
PIPTS101-10
4.
Dispense the Sample
Once all test information is entered, the PTS Reader displays:
x
ADD SAMPLE; PRESS ENTER
x
Pipette 25
μ
L of sample into all four (4) sample reservoirs
of the inserted cartridge and press
Enter
on the PTS
Reader keypad.
x
Pumps draw sample aliquots into the test channels,
thereby initiating the test.
x
Results will be obtained in approximately 15 minutes.
ENDOTOXIN TEST RESULTS
1. When the test is complete, the PTS Reader gives an audible notification that the
assay is finished.
2. At the conclusion of the test, the endotoxin measurement and the assay
acceptance criteria are displayed on the screen.
3. The PTS Reader display alternates between the following results:
x
Sample
EU/mL
x
Sample
%CV
x
Spike
EU/mL
x
Spike
%CV
x
% Spike Recovery
x
Remove
Cartridge
The PTS Reader continues to display the assay results until cartridge is removed.