Listed in the table below are various Error or Fault Messages and a description of the
corresponding cause of the error or fault.
¾
One or more LONG audible tones will precede each error message.
¾
Any fault listed here will prevent the PTS Reader from running a test.
¾
Unless it is a hardware problem, the only way to clear an error message is to turn
the PTS Reader OFF and back ON again.
MESSAGE
DESCRIPTION
“
A2D SETUP
”
Photo A/D set up did not complete
“TEMP TOO LOW”
Cartridge temperature is too low
“TEMP TOO HI”
Cartridge temperature is too high
“MOTOR NOT HOME”
Cannot run pump motor
“CLOCK FAULT”
Cannot read Real Time Clock
“Computing Error”
Divide by 0 or range error
“Charge Battery”
Batteries too weak to run tests
“Battery Fault”
Battery charging problem
“UNKNOWN”
Any unspecified system fault
“One or more test LEDs out of spec”
“Abort Test CANCEL OR ENTER”
One or more of the LEDs does not
have sufficient intensity to run the
test properly
“Sample Not Seen” flashes during
test
Sample was not detected in one or
more of the optical cells