Endosafe
®
- Portable Test System User’s Guide
Copyright 2003
Page C2
PIPTS101-10
x
Enter Sample 2 Lot #
(Sample 2 is the next well to the right of sample 1)
x
Enter Sample 3 Lot #
(Sample 3 is the next well to the right of sample 2)
x
Enter Sample 4 Lot #
(Sample 4 is the last well on the right)
x
Enter Sample 1 ID
(Selecting and scrolling with the menu key under the
sample ID Header allows for fifty (50) samples to be entered and stored)
x
Enter Sample 2 ID
x
Enter Sample 3 ID
x
Enter Sample 4 ID
x
Enter Dilution #1 Factor
x
Enter Dilution #2 Factor
x
Enter Dilution #3 Factor
x
Enter Dilution #4 Factor
x
Note:
All LAL Reagent Water (LRW) used to make product dilutions should
be tested for the absence of detectable endotoxin.
While the above information is being entered into the PTS Reader, the cartridge
is being pre-warmed.
4.
Dispense the Sample
Once all test information is entered, the PTS Reader
displays:
x
ADD SAMPLE; PRESS ENTER
x
Pipette 25
μ
L of sample into four (4) sample reservoirs
of the inserted cartridge and press
Enter
on the PTS
Reader keypad.
x
Pumps draw sample aliquots into the test channels,
thereby initiating the test.
x
Results will be obtained in approximately 15 minutes.