Appendix 1 A-B
16.2
Guidance and Manufacturer’s
declaration electromagnetic immunity
The CENTURION
TESLA 21+ LASER SHOWER
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
TESLA 21+ LASER SHOWER
laser shower should assure that it is used in such an environment.
Immunity test
Test level
IEC 60601-1-2:2014
Compliance level
Electromagnetic environment-guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 8 kV contact discharge
± 2 kV, ± 4 kV, ± 8 kV,
15 kV air discharge
± 8 kV contact discharge
± 2 kV, ± 4 kV, ± 8 kV,
15 kV air discharge
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30 %.
Electrical fast transient/ burst
IEC 61000- 4-4
± 2 kV, 100 kHz repetition
frequency
± 2 kV for power supply
lines
Mains power quality should be that of a typical commercial
or hospital environment.
Harmonic emissions IEC 61000-
3-2 (*)
+ 0,5 kV, + 1 kV
+ 0,5 kV, + 1 kV
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and voltage vari-
ations on power supply input
lines
IEC 61000-4-11
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% U
τ
; 1 cycle and
70% U
τ
; 25/30 cycle
Single phase: at 0°
0% U
τ
; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% U
τ
; 1 cycle and
70% U
τ
; 25/30 cycle
Single phase: at 0°
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the
TESLA 21+
LASER SHOWER
requires continued operation during
power mains interruption, it is recommended that the
TESLA 21+ LASER SHOWER
be powered from an
uninterruptible power supply or a battery.
0%
Uτ; 250/300 cycle
0% Uτ; 250/300 cycle
Power frequency
(50/60 Hz) magnetic field IEC
61000-4-8
30 A/m
50 Hz or 60 Hz
30 A/m
50 Hz or 60 Hz
Power frequency magnetic fields should be at levels cha-
racteristics of a typical location in a typical commercial or
hospital environment. The area shower does not contain any
sensitive parts or switch elements.
Note
: U
τ
is the a.c. mains voltage prior to application of the test level.