Appendix 1 A-B

 

 

 

16.2

 

Guidance and Manufacturer’s 

declaration electromagnetic immunity 

 

The CENTURION 

TESLA 21+ LASER SHOWER

 is intended for use in the electromagnetic environment specified below. The 

customer or the user of the 

TESLA 21+ LASER SHOWER

 laser shower should assure that it is used in such an environment.

 

Immunity test

 

Test level

 

IEC 60601-1-2:2014

 

Compliance level

 

Electromagnetic environment-guidance

 

Electrostatic discharge (ESD) 

IEC 61000-4-2

 

± 8 kV contact discharge

 

± 2 kV, ± 4 kV, ± 8 kV,

 

15 kV air discharge

 

± 8 kV contact discharge

 

± 2 kV, ± 4 kV, ± 8 kV,

 

15 kV air discharge

 

Floors should be wood, concrete or ceramic tile. If floors are 
covered with synthetic material, the relative humidity should 

be at least 30 %.

 

Electrical fast transient/ burst 
IEC 61000- 4-4

 

± 2 kV, 100 kHz repetition 
frequency

 

± 2 kV for power supply 
lines

 

Mains power quality should be that of a typical commercial 
or hospital environment.

 

Harmonic emissions IEC 61000- 
3-2 (*)

 

+ 0,5 kV, + 1 kV

 

+ 0,5 kV, + 1 kV

 

Mains power quality should be that of a typical commercial 
or hospital environment.

 

Voltage dips, short 
interruptions and voltage vari- 
ations on power supply input 

lines

 

IEC 61000-4-11

 

0% UT; 0,5 cycle

 

At 0°, 45°, 90°, 135°, 180°,

 

225°, 270° and 315°

 

0% U

τ

; 1 cycle and 

70% U

τ

; 25/30 cycle 

Single phase: at 0°

 

0% U

τ

; 0,5 cycle

 

At 0°, 45°, 90°, 135°, 180°,

 

225°, 270° and 315°

 

0% U

τ

; 1 cycle and 

70% U

τ

; 25/30 cycle 

Single phase: at 0°

 

Mains power quality should be that of a typical commercial 
or  hospital  environment.  If  the  user  of  the 

TESLA  21+ 

LASER SHOWER

 requires continued operation during  

power  mains  interruption,  it  is  recommended  that  the 

TESLA  21+  LASER  SHOWER

    be  powered  from  an 

uninterruptible power supply or a battery.

 

 

0% 

Uτ; 250/300 cycle

 

0% Uτ; 250/300 cycle

 

 

Power frequency

 

(50/60 Hz) magnetic field IEC 

61000-4-8

 

30 A/m

 

50 Hz or 60 Hz

 

30 A/m

 

50 Hz or 60 Hz

 

Power  frequency  magnetic  fields  should  be  at  levels  cha- 
racteristics  of  a  typical  location  in  a  typical  commercial  or 

hospital environment. The area shower does not contain any 
sensitive parts or switch elements.

 

Note

: U

τ

 is the a.c. mains voltage prior to application of the test level.

 

Содержание TESLA 21+

Страница 1: ...Operating Manual For Veterinary use only Last revised June 12 2020...

Страница 2: ...pson of the electron During that period in the mid 1890 s Tesla conversed often with Thompson particularly in the electrical journals At about the year 1918 Tesla apparently had a laser like apparatus...

Страница 3: ...by Coleman Czito s grandson s wife it is very possible that Tesla used the same or similar kind of apparatus to send laser pulses to the moon Working of Ruby Laser 1 A powerful lamp is wrapped around...

Страница 4: ...e action Copyrights The Operating Manual are subject to copyright law All rights are reserved Copying duplicating translating and converting into an electronic format in whole or in part are prohibite...

Страница 5: ...ns 19 12 1 Definition of symbols of the Centurion TESLA 21 LASER SHOWER 20 13 Error messages and safety information 21 14 Technical data The Centurion TESLA 21 LASER SHOWER 22 15 Appendix Definition o...

Страница 6: ...l as dose 6 LEDs 6a Two yellow LED On when laser output active Normal Operation 6b Two red LED On as error indication of LED diode operation Please contact customer service 7 red connector Connector f...

Страница 7: ...ER SHOWER casing top side 6b 6a 10 11 9 8 7 5 3 fig 2 2 shower adapter top fig comb adapter lower fig 1 See Appendix 1 2 3 PM11 8800 Li ion Battery Pack For Centurion TESLA 21 LASER SHOWER fig 3 AC ad...

Страница 8: ...nd can be used for helping wound healing and relief of pain and inflammation 2 Contraindications Theprofessionalliteratureliststhefollowingcontraindications Theyalsoapplyto the TESLA 21 LASER SHOWER A...

Страница 9: ...J LD are displayed Factory setting whendeliveringis 100mW Multi 5min 15J LD Avaluethatappearsinvertedcanbechangedwiththe and keys Onthisscreenthelaserpowerlevelcanbe altered Pressing the SET key saves...

Страница 10: ...ht alignment If you reconnect the TESLA 21 LASER SHOWER with the AC adapter make sure you observe the direction of the arrows on the plugs 3 2 Connect AC Adapter to the power supply Connect the AC ada...

Страница 11: ...hown the following initial settings The maximum output 100 mW the operation mode Multi and an energy dose of 60 J LD LD laser diode are preset when operated for the first time Following this the laser...

Страница 12: ...alue in increments of 10 Press the SET key again and the selected value is confirmed and stored The display changes showing the following information The therapy program can be changed with the and pa...

Страница 13: ...d and isflashing 3 Select the FI Bands using the or key The display now shows FI xxxxx 4 Press the arrow key for 2 3 seconds The display now shows in the first line the frequency FI autorun and in the...

Страница 14: ...to the treatment screen on the display at first as a flashing value The frequency is stored with the SET key and displayed on the screen A set value remains in the memory until it is overwritten by an...

Страница 15: ...py time for the TESLA 21 LASER SHOWER 100mW is 10 minutes The maximum energy dose rate that can be set is therefore 100 mW x 10 minutes x 60 s 60 J LD The minimum dose that can be selected is 1 J Note...

Страница 16: ...e emitted energy dose rateinJoule The endof the first line displays the mode selected The selected therapy time now counts backwards and the energy dose rate changes in line with the selected laser po...

Страница 17: ...emperature of 55 the mode is switched automatically to multi frequency mode The temperature symbol on the display will inform you page 21 During laser emission mini fans will see to an improved remova...

Страница 18: ...ion fig 9 2 In this position you can remove the adapter fig 9 3 When attaching the adapter make sure that the detents are inserted into the proper recesses Press the adapter gently against the housing...

Страница 19: ...ed please contact the Service Department The laser shower should only be used at moderate temperature 11 Safety information Please strictly observe the instructions in the operating manual and heed th...

Страница 20: ...tment Magnetic and electric fields as well as ionizing radiation may influence the functioning of the unit For this reason you should not operate near devices which generate large electromagnetic fiel...

Страница 21: ...igits 5 6 starting from the right year of manufacture 20 2020 Manufacturer s nameplate This warning sign marks the laser beam outlet of the unit Do not look into the laser beam outlet without laser pr...

Страница 22: ...and must not be disposed of in the domestic waste Distributor 12 2 Definition of symbols on AC adapter Manufacturer date only AC adapter in form YY MM Device of laser protection class II only AC Adapt...

Страница 23: ...on Laser output toohigh Whenthismessageisdisplayed theoutputemitted varies from the output selected Laser output still can continue This message may also occur due to reflection of laser light on wet...

Страница 24: ...approx 310 x 105 x 47 mm Total weight 430 g Classification Protection class II according to EN 60601 1 Application unit BF according to EN 60601 1 Without moisture protection IPX0 Laser class 3B acco...

Страница 25: ...cies Display Frequeny Multi Frequency 200 Hz to 3 5 kHz random signal derived from the quartz oscillator impulse duration max 0 25 msec Alpha Frequenz 10 Hz Fi Band Regeneration Pain Trauma Detox Infe...

Страница 26: ...s that the device meets the EMC requirements only if the accessories are used which are listed in the EC Declaration of Conformity Using other accessories may result in an increased emission of electr...

Страница 27: ...nly for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions Class B The TESLA 21 LASER SHOWER laser s...

Страница 28: ...3 2 0 5 kV 1 kV 0 5 kV 1 kV Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage vari ations on power supply input lines IEC...

Страница 29: ...MHz 5 7GHz 9 28V m NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and ref...

Страница 30: ...ment parts Freight excluded Any repairs or service work done by an unauthorized person or company will void the Warranty immediately An Extended Warranty Package is available to cover any problems tha...

Страница 31: ...VE EDISON N J 08837 PHONE 1 732 417 0500 PHONE 1 800 387 8326 Email info centurionsystems com Delivered by ELEC WESTERN MEDICAL DEVICES LTD 1015 Matheson Blvd East Suite 8 Mississauga On L4W 3A4 PHONE...

Отзывы: