16.
Notes on Electromagnetic Compatibility
Medical electronic devices are subject to special precautions regarding
their EMC and must be installed and operated in accordance with the EMC
instructions in the accompanying documents.
Warning
: Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used no closer
than 30 cm (12 inches) to any part of the laser shower including the cable
and plug-in power supply. Otherwise, degradation of the performance of
this device could result.
Warning: The manufacturer warrants that the device meets the EMC
requirements only if the accessories are used which are listed in the EC
Declaration of Conformity.
Using other accessories may result in an
increased emission of electromagnetic interference or reduced immunity
against electromagnetic interference.
Warning
: The device may not be placed directly next to or stacked with
other devices. If such an arrangement is nonetheless necessary, the device
must be observed to make sure that it operates properly in this
arrangement.
Higher level of electromagnetic disturbance could degrade or abandon the
laser output temporarily.
While respecting the hints to EMC and guidance to EMC of this user manual,
no restriction is expected to the essential performance of the medical
device during its lifetime of 7 years.