CenterVue
DRS Operating Manual
Pg. 16 of 59
Figure 9 - Auto-focus in progress
Figure 10 - Image processing and storage in
progress
9.5 Manual alignment
If, for some reasons, the automated eye search fails, a message as in 11 shows up. Reasons for
failure include: improper patient positioning, patient not fixating steadily, patient moving the head
and / or the gaze during eye search.
If any of the above situations applies, try to fix the cause (re-position the patient,
instruct her/him to steadily fixate at the target and/or to not move) before proceeding
with the manual alignment.
Use the UP / DOWN / LEFT / RIGHT buttons to adjust the optical head position and align the
patient
’s eye to the instrument front lens. As soon as the eye is detected the automated procedure
will resume.
Figure 11 - Failure of automatic eye search
The result of the overall acquisition process is one or more retinal images as the one shown below.