17. CLEANING
This paragraph explains how to clean the system. The chin rest and the forehead rest shall be
disinfected between one patient and the next. The disinfecting solution should be applied using a
wipe, taking care not to sprinkle parts not belonging to the patient rest.
After each examination, between one patient and the next, the forehead rest and the chin rest shall
be cleaned using isopropyl alcohol or another similar disinfectant solution, not containing oxidizing
agents.
Extreme care should be given to always protect the front lens.
The front lens should be cleaned by using a small hand pump air blower, to blow away dust. Only if
really needed, for instance due to the presence of a fingerprint, the front lens should be cleaned
using photographic cleaning paper and a suitable lens cleaning fluid.
The touch screen panel should be cleaned only with a cloth damped in water.
Do not use alcohol or detergents to clean the touch screen as these may damage the
film covering the panel
When cleaning the rest of device, the device must be powered off, and the power cord shall be
disconnected from mains. If needed, the external covers of the unit can be cleaned by means of a
slightly damp cloth.