51
13. Insert the vessel stand in the instrument cavity with the standoffs pointing downward. Ensure that the vessel stand
is flat against the bottom of the cavity and that the cutout in the stand is positioned over the opening in the cavity
liner.
14. Place 100mL of water and a stirring bar into a 100mL round bottom flask.
15. Place the flask into the cavity so that it is positioned on top of the vessel stand.
16. Install the open vessel attenuator assembly with the large slot positioned toward the back of the instrument.
17. Rotate the attenuator assembly clockwise until the attenuator locks into position.
18. Insert the fiber optic probe into the vessel.
19. Press the START/PAUSE key.
Note:
If the following screen appears, ensure that the Fiber Optic option is assembled correctly. Also ensure that Fiber
Optic is selected and calibration has been entered.
20. Once the START/PAUSE key is pressed, the following message will appear in the bottom line of the instrument
Screen - “Record start temperature in 10 seconds.” Once the countdown is complete, the bottom line will read
“Applying microwaves for 60 seconds”. After 60 seconds “Capturing end temperature” will appear. “Power Test
Complete” will appear along with the correct microwave wattage that is being applied.
Note:
The Discover SP-D Clinical system is designed to apply 300 watts. If the current wattage is greater than or less
than ten percent, the system is within its operating limits. If the system is not within ten percent, calibrate the instrument.