9
NOTE:
The regulator has an arrow printed on it to show the direction that the air must flow. Set up the regulator according to
the arrow.
d. Place the second length of tubing into the opposite end of the regulator.
e. Connect the opposite end of the tubing to the air source.
f. To change the amount of air flow from the regulator:
Pull the knob on the regulator outward.
Rotate the knob clockwise to increase the amount of air flow.
Rotate the knob counterclockwise to decrease the amount of air flow.
The minimum compressed gas requirement is 25 PSI (20L/min flow).
The maximum flow rate is 60 PSI.
Note:
The air flow should remain on at all times when a reaction is being performed.
7. Check the cavity for debris.
8. Using the power switch located on the left side of the Discover, turn the instrument on.
9. The ActiVent (pressure device) will return to the home position once the Login screen appears. Remove the 10mL
shipping device.
Note:
Do not discard the 10mL shipping device. It will be required if returning the system to CEM Corporation for service.
Computer Installation (optional)
1. Connect one end of the Ethernet cable to one of the Ethernet ports located at the rear of the Discover.
2. Connect the opposite end of the Ethernet cable extending from the Discover to the computer.
3. Ensure that the computer is turned on and the desktop is visible. Select the Synergy icon to activate the Synergy
software. If the Synergy icon is not visible on the desktop, insert the Synergy CD and follow the on-screen
instructions.
Note:
If this is the first time the software has been activated on the computer, the user will be prompted to create a new
account.