44
Note:
Depending on the viscosity of the sample, a reaction mixture can be prepared and the above procedure
can be used to determine if the sample is adequately stirring.
3. Log in to the Discover SP-D Clinical as a user with Administrative privileges.
4. Press the HOME key to display the “Discover Home” screen.
5. Press the “+/-“ keys to highlight “Administrative Functions.”
6. Press the ENTER key.
7. Press the right ARROW key one time to display the “Administration” screen.
WARNING
Proper precautions must be taken to avoid contact with solvents or solvent vapors. Protective gear should
be worn as outlined in the user’s safety program for hazardous materials and the reagent manufacturer’s
material safety data sheet. Refer to these guidelines for proper handling and disposal of reagents.
8. Press the ENTER key.
9. Press the right ARROW key for (4) times to display the “Maintenance 5” screen.
10. Press the “+/-” keys to highlight “Test Stirring.”
11. Press the ENTER key. The stirring speed will be displayed at the bottom of the screen.
12. Press the “+/-“ keys to change the stirring speed to High, Medium, Low or Off.
Note:
If stirring does not occur, try using a larger/stronger stir bar.
Note:
The prepared sample can be used to determine adequate stirring speed for each sample.
Note:
If stirring does not appear to be working properly, contact CEM Service at
800-726-5551
or
704-821-7015
or email
.
Содержание Discover SP-D
Страница 1: ...i Operation Manual 600857 Rev 1 8 17 ...