26
4. Press the ENTER key. A new screen will appear.
5. Use the ARROW and “+/-” keys to highlight the “HOME” button.
6. Press the ENTER key. The APM will return to the HOME position.
CAUTION
Improper use of the APM Recovery screen could result in destruction of the APM, making the warranty null and
void.
The pressure assembly needs to be in the highest position before the APM is moved to the back of the
instrument.
7. If the APM does not return to the home position:
a. Follow steps 3-5 above.
b. Use the ARROW and “+/-” keys to highlight “Z UP.” Press the ENTER key. If the motor can be
heard, continue to select “Z Up” until the motor no longer sounds like it is turning. This will raise
the ActiVent to the highest position.
c. Use the ARROW and “+/-” keys to highlight “APM Back.”
d. Press the ENTER key. If the APM appears to be moving backward, select “APM Back” until the
ActiVent no longer moves backward. The ActiVent should be all the way back, away from the attenuator.
e. Use the ARROW and “+/-” keys to highlight the “HOME” button.
f. Press the ENTER key. The APM should now return to the HOME position.
System Setup
The “System Setup” function consists of five (5) “pages” that allow the user to calibrate system settings, set method
defaults, clean the ActiVent feature, and view system information.
NOTE:
“System Setup” can only be viewed by Standard and Administrator users.
To enter the System Setup screen:
1. Press the HOME key until the “Discover Home” screen appears.
2. Use the “+/-” keys to highlight “System Setup”.
3. Press the ENTER key. A new screen will appear.