27
User Setup
The “User Setup” page is the 1
st
page that appears after selecting “System Setup” from the Discover home screen. This
page allows the user to set up method defaults and general operating parameters for the Discover SP-D Clinical system.
Temperature Type
“Temperature Type” allows the user to switch between the Infrared and Fiber Optic temperature control.
1. Using the “+/-” keys, highlight “Temperature Type.”
2. Press the ENTER key.
3. Using the “+/-” keys, highlight the applicable temperature control to be utilized (infrared or fiber optic).
NOTE:
To utilize the fiber optic control, the fiber optic temperature control option must be connected on the side of the
instrument. The part number for the fiber optic assembly is 541166.
4. Press the ENTER key. The “User Setup” screen will appear again.
Release limits
Temperature and pressure release limits are the temperature and pressure at which the ActiVent will
automatically open and release the vessel after cooling. The recommended default temperature is 60°C. The
recommended default pressure is 2.8 Bar (40 PSI).
NOTE:
The default temperature should be lowered if a solvent with a low boiling point is used.
1. Using the “+/-” keys, highlight “Release Limits.”
2. Press the ENTER key.
3. Use the “+/-” keys to increase and decrease the release limits. Use the ARROW keys to toggle between
“Temperature” and “Pressure.”
4. Once the parameters have been entered, press the ENTER key. The “User Setup” screen will appear.
Содержание Discover SP-D
Страница 1: ...i Operation Manual 600857 Rev 1 8 17 ...