9.2 Blood Pressure Module
Measuring method
auscultatory method (Korotokov), oscillometric; for resting BP,
the results from both measurements are compared for
plau sibility
Measuring range
systolic pressure: 40 to 280 mmHg
diastolic pressure: 40 to 280 mmHg
pulse rate:
35 to 230 bpm
Measurement error, systematic
systolic pressure: +/– 3 mmHg
diastolic pressure: +/– 3 mmHg
(temperature: +15 to +25 °C)
Standard deviation (clinical trial)
systolic/diastolic pressure: 7 mmHg (max.)
Inflation pressure
300 mmHg max.; during inflation the inflation pressure
automatically adapts to patient’s BP
Inflation rate
between approx. 6 seconds (to 140 mmHg) and approx.
18 seconds (to 300 mmHg)
Max. cuff pressure
300 mmHg
Cuff deflation method
pulse-dependent deflation rate
approx. 3 mmHg/beat or approx. 3 mmHg/s
Calibration
calibration with external pressure meter
Artifact rejection
automatic artifact rejection