User Manual: xr100+_eng
27
Cardioline S.p.A.
Via Linz,151
38121 Trento (TN)
www.cardioline.com
Sales office:
Via F.lli Bronzetti, 8 20129 Milano (MI) Italy tel.+39 02 94750470
fax +39 02 94750471
xr100+
EKG Type
The selected EKG Type determines the communication
method with the ECG recorder, PC-based ECG system, etc.
To prevent an accidental change of this setting, the menu is
protected with a password.
Using the arrow keys, enter 003 and confirm the entry with
Select
.
Figure 7 – 28: Entering the EKG Type password
All ergometers support the following communication
modes:
•
Analog with pulse
Remote start mode; before advancing to the next load
level, the ergometer generates a control pulse and
sends the corresponding data via the interface.
•
Analog / Digital
An analog voltage controls the load – blood pressure
measurements can be initiated with digital commands.
•
Digital
(default)
The communication with the ergo meter is entirely
controlled with digital commands.
•
Analog IN-OUT
The entire communication (load control and BP mea-
surements) is controlled with analog signals.
No digital data will be sent.
Select the communication mode and confirm with
Select
.
Figure 7 – 29: Selecting the ergometer communication mode
Note
• The EKG Type needs to be selected only when the ergometer
is connected to an ECG unit. The selection is part of the
installation procedure.
• The “Analog/Digital” and “Digital” communication is only
possible when PC Mode is selected from the main menu or
when this is the default mode.