28
User Manual: xr100+_eng
Cardioline S.p.A.
Via Linz,151
38121 Trento (TN)
www.cardioline.com
Sales office:
Via F.lli Bronzetti, 8 20129 Milano (MI) Italy tel.+39 02 94750470
fax +39 02 94750471
xr100+
RPM
Here you determine the RPM limits. When these limits
are exceeded, the LEDs for high or low speed (RPM) will
illuminate.
Select the value to change (Min. or Max.) and confirm with
Select
.
Using the arrow keys, change the value and save the new
value with
Select
.
Figure 7 – 30: Setting the RPM limit values
Note
The limits selected in this menu only apply to the load range
between 6 and 150 watts. At higher loads the RPM limits auto
matically adapt to the respective loads:
Load (watts)
Green RPM range (1/min)
6 – 150
54 – 64 (
adjustable
)
151 – 250
58 – 65
251 – 350
68 – 75
351 – 450
78 – 85
451 – 550
88 – 95
551 – 650
98 – 105
651 – 750
108 – 115
751 – 850
118 – 125
851 – 950
> 125
951 – 999
> 130
Pulse Display
The pulse readout on the display can be turned off.
Figure 7 – 31: Setting the pulse readout