Percentage calculation of the electrode status is ba-
sed both on asymmetry potentials and on slopes (gra-
dient of the pH characteristic) of the electrode. The
electrode condition is then determined by the worst
result. The admissible characteristic data of the pH
electrodes are as follows:
Asymmetry potential of the electrode at 25
°
C in mV
Max. admissible range:
+30...-30
mV
Optimum:
0
mV
Soiling of the electrode can have a major adverse ef-
fect on asymmetry.
[SL] = Slope of the electrode signal at 25
°
C in mV/pH
Admissible range:
-62...-50
mV/pH
Optimum:
-59,2
mV/pH
1-point calibration is always based on the optimum
theoretical value of –59.2mV/pH. The slope is not
shown in the menu.
For 2-point calibration, the slope is specified for the
entire measuring range.
For 3-point calibration, the slope is first displayed for
the lower, acid measuring range [SL. 1] and then for
the upper, basic range [SL. 2].