16
Performing calibration
If you have not already done so, select the measuring
mode 'pH' (see Configuration) and ensure that (de-
pending on your requirements), 1-point, 2-point or 3-
point calibration is active in the configuration setting
Carefully remove the protective cap from the electro-
de.
Clean the electrode using deionised water and
dry carefully.
9.1 Starting calibration: Press [Cal] key.
‘CAL’ appears momentarily in the display.
You can terminate calibration at any time by pressing
the [Meas] key or [Hold] key. In this event, the previous
calibration setting is retained.
9.2 Selecting buffer solutions
‘buFF’ is shown in the main display.
You can choose between the various buffer series using
the
[
▲
] [
▼
]
keys.
You then confirm the selected buffer set by pressing
the [Store] key.
The following inputs are possible:
Tintometer standard series
Values at 25
°
C: 4.00pH, 7.00pH, 10.00pH
Store
+
Store