DIN 19266 series
Values at 25
°
C: 1.68pH(A), 4.01pH(C), 6.87pH(D),
9.18pH(F),12.45pH(G)
Manual buffer setting
If you want to use buffers other than those in the Tin-
tometer standard or DIN series, you can choose ma-
nual buffer adjustment here.
9.3 Calibration point 1: 'CAL. 1'
Insert the electrode and the temperature sensor (if
present) in buffer solution 1 (e.g. Tintometer standard
buffer solution pH 7.00) and swirl carefully.
In the case of manual buffer setting, [Pt.1] appears in
the secondary display.
You enter the pH value of the buffer solution using the
[
▲
] [
▼
]
keys.
You then confirm the input by pressing the [Store] key.
When Tintometer standard and DIN series solutions
are used, the system automatically detects the pH
value of the solution.
[CAL.1] appears in the secondary display until the va-
lue is stable; the detected value is then automatically
stored .
You are then requested to perform the next calibration
step.
Store
+
Store