5
The permanent protective plating on the unit serves
not just as impact protection but also as a storage
compartment for the electrode and as an anti-slip de-
vice when the meter is used in an upright position.
Microprocessor technology and 4 x AA 1,5 V batte-
ries allow unit operation without battery change for
15,000 hours or more.
A "Low Bat” indicator in the display shows the user
when it is time to change the batteries.
The required temperature compensation for the sy-
stem is performed automatically (ATC) following con-
nection of a temperature sensor or via manual input
via the touch-sensitive film keypad. The unit automati-
cally identifies Pt1000 and NTC 30 kOhm sensors.
The AUTO HOLD feature enables the user to identify
stable results and call them up on the display.
The MEAS function is the standard continuous measu-
ring mode for pH levels or redox voltages.
pH calibration of the system is via an integrated auto-
matic buffer recognition feature.
The temperature characteristics of the standard
buffer solutions and the reference buffer solutions
specified in DIN 19266 are stored in the unit and
are used for automatic calibration of the system when
selected.
It is additionally possible to enter some of the buffer
data into the unit manually.
Calibration is performed optionally at 3 points, 2 points
or 1 point.
A sensor evaluation is automatically shown in the dis-
play (or can be called up via the menu) after each ca-
libration.