8. Configuration of the unit
To configure the unit, press the [Menu] key. The first
menu item appears.
If you press the [Menu] key again, the display jumps to
the next menu item.
You can set the various parameters using the
[
▲
] [
▼
]
keys.
Press the [Store] key to store and conclude configura-
tion.
The unit then switches back to measuring mode.
You can terminate the configuration process by pres-
sing the [Meas] or [Hold] key.
Settings made up to this point are not stored.
The unit retains the previous settings.
8.1 Input: selection of measuring
function [pH] – [mV] – [Rel mV]
Select the desired measuring function using the
[
▲
] [
▼
]
keys. The arrow in the display shows the selected func-
tion.
• pH:
pH measurement with pH electrode
• mV:
redox measurement with redox electrode
• Rel mV:
redox measurement with redox electrode.
The displayed value is converted to the
normal hydrogen system in temperature-
dependent mode (DIN 38404).
Menu
Store
+
+
Hold
Meas
Store
Meas
Hold