D-46
Appendix D
Scanner
®
2000 microEFM
Bit Definitions—Alarms and Diagnostics
Bit
Flow Run
Alarm High
Flow Run
Alarm Low
Diagnostic 1
Diagnostic 2
Diagnostic 3
Diagnostic 4
2
FRA3 High
FRA3 Low
PT Override
PT Low
MVT User
Parameter
CRC fail
Reserved
1
FRA2 High
FRA2 Low
A1 Override
A1 Low
MVT Factory
Parameter
CRC fail
Device Seal
0
FRA1 High
FRA1 Low
A2 Override
A2 Low
MVT Not
Present
Ext. Switch
The Scanner 2000 produces low, high and fail conditions for the inputs (not the flow alarms) in accordance
with the following table.
Status
Range Check
Low
Lower Range Limit - 20% of span
Fail Low
Lower Range Limit - 500% of span
High
Upper Range Limit + 20% of span
Fail High
Upper Range Limit + 500% of span
Units of Measurement
Value
Units
Scale
Offset
101
GAL - BASE
1.000000000000
0.00
102
BBL
0.023809523810
0.00
103
M3
0.003785411780
0.00
104
LIT
3.785411784000
0.00
105
CF
0.133680555560
0.00
106
ACF
0.133680555560
0.00
107
ACM
0.003785411780
0.00
201
MCF - BASE
1.000000000
0.00
202
SCF
1000.000000000
0.00
203
M3
28.316846592
0.00
204
GAL
7480.519480271
0.00
205
BBL
178.107606673
0.00
207
LIT
28316.846592000
0.00
301
PSIG - BASE
1.00000000
0.00
302
Pa
6894.75729317
0.00
303
Kpa
6.89475729317
0.00
304
Mpa
0.00689475729317
0.00
305
Bar
0.06894757
0.00
306
In H2O
27.70500000
0.00
401
In H2O - BASE
1.000000000000
0.00
402
Pa
248.641080600000
0.00
403
Kpa
0.248641080600
0.00
Содержание Scanner 2000 microEFM
Страница 1: ...NUFLO Scanner 2000 microEFM Hardware User Manual Manual No 9A 30165023 Rev 10 ...
Страница 26: ...26 Section 1 Scanner 2000 microEFM ...
Страница 104: ...A 18 Appendix A Scanner 2000 microEFM ...
Страница 106: ...B 2 Appendix B Scanner 2000 microEFM ...
Страница 188: ...E 10 Appendix D Scanner 2000 microEFM ...
Страница 190: ...R U S S I A 603 5569 0501 ms kl c a m com ...