![CAI 700M Скачать руководство пользователя страница 83](http://html1.mh-extra.com/html/cai/700m/700m_operators-manual_3331460083.webp)
700M HFID
6-2-15
83
Reset Offsets and Gains
→
→
→
→
Pressing
from the Reset Factory Settings menu will prompt the operator to confirm
resetting Offsets and Gains for all ranges. Pressing
(Yes) from this screen resets the Offset
and Gain factors to factory default settings (0 and 1 respectively) and brings you to this
confirmation screen:
•
Offset and Gain factors are created when the analyzer is zeroed and spanned.
•
If the Offsets and Gains are reset, the analyzer must be zeroed and spanned again before
use.
•
All recorded deviations will be set to zero.
If you press
(No) from the Reset Offsets and Gains screen, the analyzer will return to the
Reset Factory Settings menu without resetting the Offsets and Gains.