The following output signals can be programmed by the operator using the Output
Assignments screen:
RealTime
: In either THC or CH
4
mode the concentration’s output will be a live reading.
CH
4
:
In CH
4
mode the output concentration will be a live reading. In THC/CH
4
/NMHC mode
the output will be a read and hold.
THC
: In THC mode the concentration’s output will be a live reading. In THC/CH
4
/NMHC mode
the output will be a read and hold.
NMHC
: The calculated concentration’s output will be updated after each complete cycle in
THC/CH
4
/NMHC mode.
SamplePres:
Sample pressure (psig).
AirPres
: Air pressure (psig).
FuelPres
: Fuel Pressure.
AirInjPres
: Air Inject Pressure.
FuelInjPres
: Fuel Inject Pressure.
FilterTemp
: Filter temperature (
°
C).
BurnerTemp
: Burner temperature (
°
C).
OvenTemp
: Oven temperature (
°
C).
CutterTemp
: Cutter temperature (
°
C).
PumpTemp:
Internal pump temperature (
°
C).
SampleEPC
: % of voltage supplied to the Sample electronic proportioning control valve.
AirEPC:
% of voltage supplied to the Air electronic proportioning control valve.
FuelEPC:
% of voltage supplied to the Fuel electronic proportioning control valve.
AirInjEPC:
% of voltage supplied to the Air inject electronic proportioning control valve.
FuelInjEPC:
% of voltage supplied to the Fuel inject electronic proportioning control valve.