Output Adjustments
→
→
→
→
→
The Output Adjustments screen allows the operator to set the output to either mA
or voltage and calibrate the outputs to exactly match the results obtained on a PLC
or other remote data-recording device.
Using the Output Adjustments screen, the
operator can force the analog outputs to 0%, 50% or 100% of full scale and back to
Measure. For information on analog output connections see
From the Programmable Analogs menu press
to access the Output Adjustments
screen.
•
Output Type
can be set for mA or 1, 5 or 10 VDC.
•
Output AO-1
refers to analog output 1.
•
% FS
is used to toggle between measurement, 0%, 50% and 100% Full Scale.
•
Offset
(zero) is used to adjust the output at 0%.
•
Gain
(span) is used to adjust the output at 100%.
To select the
Output Type
press
to open the highlighted field. Use the Up/Down
arrows
to make your selection. Press
again to close the field.