Calibration Via Pump/Valves
→
→
→
→
The use of Cal via Pump/Valves depends upon how calibration gases are being introduced
into the analyzer – via a sample pump or internal valves (if equipped with internal valve
option).
The existing setting (Valves in the example) is shown at the top right of the menu. To
access the Cal via Pump/Valves menu, press
from the Calibration Setup menu.
Press
(Cal via Pump) to keep the analyzer’s internal sample pump on and keep the valves
closed during calibration. You will return to the Calibration Setup Menu. Please note that the
Calibration Setup menu will display the current settings on the right side of the screen after the
ellipsis
(…).
Example:
Cal Pump/Valves . . . Pump.
NOTE:
If the analyzer is equipped with a pump, to prevent damage to the pump do not
pressurize the sample inlet.
Press
(Cal via Valves) to activate the appropriate calibration valve and keep the internal
sample pump turned off during calibration. Keeping the sample pump turned off while the valves
are activated will prevent sample from being mixed with calibration gas. You will return to the
Calibration Setup Menu. Please note that the Calibration Setup menu shows the current settings
on the right side of the screen after the ellipsis
(…).
Example:
Cal Pump/Valves . . . Valves.