![CAI 700M Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/cai/700m/700m_operators-manual_3331460059.webp)
700M HFID
6-2-15
59
Initiate Sequenced Check
→
→
→
→
Initiate Sequenced Check is a useful tool for setting up Auto Calibration.
From the Auto
Calibration Menu screen, pressing
initiates a sequenced calibration
check.
Rather than
initiating a calibration, it checks the validity of your most current calibration.
A sequenced calibration check performs all of the steps of a
exception
of the zero and span
calibrations.
It does not set new offsets, gains or any alarms.