![CAI 700M Скачать руководство пользователя страница 168](http://html1.mh-extra.com/html/cai/700m/700m_operators-manual_3331460168.webp)
700M HFID
6-2-15
168
Control Commands
SARA: Auto Range Off
Command
Response
Description
_SARA_K0
_SARA_s
Set auto range Off.
SARE: Auto Range On
Command
Response
Description
_SARE_K0
_SARE_s
Set auto range On.
SATK: Start Automatic Calibration
Command
Response
Description
_SATK_K0
_SATK_s
Start automatic calibration of all available ranges.
_ SATK_K0_Mn
_SATK_s
Start automatic calibration of range ‘ n ‘.
SCH4: Sets CH4 Mode
Command
Response
Description
_SCH4_K0
_SCH4_s
Sets the measurement mode to CH4 only
SEGA: Open Valve for Span Gas Calibration
Command
Response
Description
_SEGA_K0
_SNGA_s
Sets to Span Calibration mode.
_ SEGA_K0_Mn
_SNGA_s
Open to Span Calibration mode Range n.
SEKA: Saves Measured Value as New Span Value
Command
Response
Description
_SEKA_K0
_SEKA_s
Saves measured value of actual range as gain if
Span mode is set.