700M HFID
6-2-15
55
Automated Calibration
→
→
→
An automated calibration is a timed zero calibration immediately followed by a timed span
calibration.
The Automated Calibration menu offers two choices: Sequenced Calibration and
Sequenced Check of the existing calibration.
The Automated Calibration menu is accessed by pressing
from the Calibration menu.
Sequenced means that the flow times of both zero and span gases are controlled using a timer.
See
located in the Calibration Setup menu.
NOTES:
•
An automated calibration should not be attempted before manual zero and span
calibrations have been successfully performed.
•
If a manually initiated sequenced calibration or sequenced calibration check is selected, it
will apply only to the range that is currently in use. (Each additional range must be
calibrated separately). This also applies if the analyzer is in auto range.