03H Read Floating Point
Modbus Commands Use TCP/IP Port 502. (Do not change TCP/IP port from 7700.)
Register Numbers and Descriptions
Register Number
Contents IEEE
40001
UNDILUTED Real Time Concentration = Diluted Conc* Dil. Ratio / 10000
40003
DILUTED Real Time Concentration
40005
Concentration before Linearization & Zero / Span Corrections
40007
Raw Detector Volts
40009
CH
4
Concentration - Switching Mode
40011
NMHC Concentration - Switching Mode
40013
THC Concentration
40025
Current Range Full Scale Concentration
40031
Sample Pressure
40033
Air Pressure
40035
Fuel Pressure
40037
Air Inject Pressure
40039
Fuel Inject Pressure
40041
Filter Temp
40043
Burner Temp
40045
Oven Temp
40047
Cutter Temp
40049
Pump Temp
40051
Sample EPC Coil V
40053
Air EPC Coil V
40055
Fuel EPC Coil V
40057
Air Inject EPC Coil V
40059
Fuel Inject EPC Coil V
40061
Range 1 Offset