Feature
Description
Time/Date
Set up the time and date information printed on Battery
Labels, Battery Service Reports and Event Log.
Analyzers with an option board have battery backup RAM
that retains dates and times on power interruption or when
the analyzer is turned off.
Default: current date and time.
Battery Startup
Specify if the analyzer will ask for the Battery ID and/or
C-Code when a battery is inserted. Specifying the Battery
ID and C-Code prevents the use of wrong settings.
Default: No for Battery ID and Yes for Battery C-Code.
Program Settings
Edit program settings. Refer to Editing Program
Settings.
Company Name
Set up the company name for printing on Battery Labels
and Battery Service Reports. You can enter up 20
characters/numbers.
Default: CADEX ELECTRONICS.
Calibration
(Shortcut: Fn-8)
Correct variations of component tolerances due to aging.
Calibrating the voltage once a year maintains
performance specifications, especially for Li-Ion batteries.
Station Calibration compensates for voltage loss from
cables and electrical contacts.
Output Device
Connect the analyzer to a printer, label printer or
BatteryShop.
Sound Control
Turn on or off the audible signals for Key Press events
and Alert events.
Default: Yes for both Key Press events and Alert events.
Voltage Display
Set voltage display mode to Volts/Cell or Volts.
Default: Volts/Cell.