Possible Reason(s) / Solution(s)
Resistance test
OHMTEST
(Code 27)
A manually selected or programmed resistance test for the
battery is in progress. Program completes in 5 seconds.
Resistance test
OHMTEST
(Code 28)
An automatic resistance test under the Life Cycling
program is in progress. The program should complete in 5
seconds.
Resistance thrs.
Exceeded
FAIL 135
The battery resistance has exceeded the set resistance
threshold in OhmTest. Program has terminated.
Resting XX:XX
RESTING
(Code 19)
Station is in rest period as specified in the Custom
Program.
Resting XX:XX
RESTING
(Code 21)
Station is in an automatic rest period specified in the
Charge program for NiMH batteries. The procedure is
automatic if temperature sensing for the adapter is not
enabled.
S
Setting up
calibration
CAL WAIT
Station is preparing for a calibration process.
Shorted or
reversed
SHORT/REV
Battery voltage is too low on insertion. Battery may be
connected backwards in the adapter. Ensure battery leads
are connected properly.
Battery may be fully discharged. Recharge battery in its
original charger before placing it in the analyzer.
Battery may have shorted cells. Discard the battery.
SM-Bus adapter
inserted
SM ADP IN
SMBus Adapter is detected on a station.
Smart battery
inserted
SM BAT IN
Smart battery is detected on a SMBus Adapter.
Soft battery
FAIL 128
Battery voltage is rising quickly during charging, even after
code 118 (Soft battery step down). Processing has
terminated.
Battery may be overcharged. Discharge the battery for 10
minutes, then charge again.
Battery may be a high capacity type battery. Reduce the
Charge rate.
Battery is new. Use the Prime program to prepare the
battery for use.
Check that correct contacts are used.