54
C7000 & 7000ER Battery Analyzer V3.5
Detail
Global
Possible Reason(s) / Solution(s)
Batshop mode wait
BATSHOP
Battery is detected on an analyzer connected to
BatteryShop. Global message should be displayed. If not,
check the C-Code in BatteryShop and ensure that all the
settings are correct for the battery.
Battery inserted
INSERTED
Battery is detected on an adapter.
Battery over temp
OVERHEAT
Battery is overheated during charge. Station will go to
Resting 00:00 (code 19) until the battery cools.
If using a non OEM battery, the battery thermistor may be
incorrect for the adapter. Turn off temperature sensing in
the C-Code. Note: For safety purposes, this step is not
possible if the C-Code chemistry is Li-Ion. Contact
Cadex to upgrade the adapter for your particular
battery.
Battery removed
REMOVED
(Code 17)
Battery is removed when the program completed. This
message appears instantaneously when the battery is
removed.
Battery removed
REMOVED
(Code 26)
A 5 second delay after the battery is removed when the
program completed.
Battery shorted
FAIL 122
Battery is not accepting charge in the initial charge cycle
(the voltage remains low). Check that the battery voltage is
above 0.30V/Cell. The Charge rate may have to be
lowered. If all else fails, discard the battery.
Battery too cold
COLD WAIT
Battery temperature is below set value when inserted. The
battery will automatically start processing when
temperature reaches minimum value set in the C-Code.
Warm the battery to room temperature.
For non OEM battery, turn off temperature sensing in the
C-Code. (see code 14, Battery over temp.)
Battery too hot
HOT WAIT
Battery temperature is too hot when inserted. Cool the
battery or use the analyzer in a cooler room.
For non OEM battery, turn off temperature sensing. (see
code 14, Battery over temp.)