Possible Reason(s) / Solution(s)
Soft battery step
down
WARNING
Battery voltage rising too quickly while charging. Analyzer
is cutting the Charge rate in half and will attempt to
complete service at lower current rate. Wait until
processing is complete. If voltage rises again, code 118
(Soft battery) will appear and the program terminated.
Charge rate in the Extended C-Code may have to be
reduced. If the battery is new or in storage, use the Prime
program.
Start battery process
START
Battery service has started.
Station calibrating
CALIBRATE
Station is in calibration. Process takes about 10 seconds.
Station off line
OFFLINE
Station is not reading the adapter. Remove the adapter
and restart the analyzer. Delete the C-Code that was
selected for the battery. Make sure all your analyzers have
the same firmware version number. Press Alt-0 to reset the
system.
Suspended Process
Aborted
INTERRUPT
Battery is removed during service for more than 5 seconds.
Program terminated.
For non OEM battery, turn off temperature sensing.
T
Target capacity not
met
WARNING
Final capacity of the battery is below the Target Capacity.
The analyzer will attempt to improve the capacity by
reconditioning the battery. Wait until processing is
complete. The Auto and Prime programs will attempt to
correct this warning. If corrected, code 195 (Cap. improved
to target) will appear. Check that the mAh rating of the
battery matches the C-Code mAh setting. Battery may be
old and operating time will be less than manufacturer’s
specification.
Temperature hold -
Cooling
SYS-TEMP
This is normal. All cycles are briefly suspended due to high
temperature inside the analyzer. Service will resume in a
few minutes or after the board has cooled. This may occur
a number of times per hour. If it reoccurs continually, move
the analyzer to a cooler room.