Messages and Warnings 12
59
Detail
Global
Possible Reason(s) / Solution(s)
O
Over voltage
FAIL 120
Battery voltage is too high on insertion. Discharge the
battery in its original equipment. Check the electrolyte. If all
fails, discard the battery.
P
PIC communications
error
PIC ERROR
An internal communication error has occurred inside the
analyzer. Turn off and turn on the analyzer or press Alt-0 to
reset the system.
Plateau timeout
WARNING
Battery was fully charged before full-charge conditions
were met. For SLA and Li-Ion batteries only. The End of
Charge setting in the C-Code may have to be increased.
Battery may be old and operating time will be less than
manufacturer’s specifications.
Power on
POWER ON
Power is detected inside the analyzer.
Process completed
FINISHED
Battery service completed.
Process suspended
INTERRUPT
Battery is removed during service. Re-insert battery within
5 seconds to resume service.
Program fault
detected
WARNING
The Custom Program has produced an error and the Next
statement is not processed. The program is terminated.
Check the Custom Program statements.
R
Ready
READY
Battery is ready. Faults or warnings (if any) were corrected.
Remove battery and use as normal.
Recondition timeout
WARNING
Recondition time has exceeded the expected value for the
battery. Program proceeds to the next cycle. The
Recondition Discharge in the Extended C-Code may have
to be raised.
The battery may be intrinsically safe, preventing deep
discharge. Use I/S settings for the battery. Refer to
Intrinsically Safe Settings.
Cells are mismatched, or if battery is new, use the Prime
program.
Reconditioning
RECOND
Battery is currently being reconditioned.