2.1.1 Power supply terminals
To fasten the cable ends, isolate or strip off 7 to 10 mm of the sheathing. Now insert an insulated
screw driver or a corresponding tool into the WAGO connector as shown in the figure and clamp
the cable end down. Then remove the screw driver again.
These connections permit cables with cross-sections of 0.08 mm
2
up to 2.5 mm
2
to be clamped
down.
The existing plastic sleeves are provided for smaller cable cross-sections < 0.5 mm
2
. For all
bigger cross-sections these sleeves have to be removed.
Version AC [230/115 V 50/60 Hz]
PIN 1 - PHASE AC
PIN 2 - GND
PIN 3 - NEUTRAL AC
2.1.2 Clamping the cable ends to the connector