8 | Preparations for a determination
Büchi Labortechnik AG
72/106
Operation Manual K-365 Dist Line
8.7.3
Changing the name of a user
Specifications on the display:
u
Navigate to the specified submenu.
User Administration
u
Select the user that you wish to edit.
...
...
u
Tap the specified function on the function bar.
EDIT
u
Enter a new name for the user.
...
u
Tap the specified function on the function bar.
ð
The new name is saved.
SAVE
8.7.4
Selecting a user
Specifications on the display:
u
Navigate to the specified submenu.
User Administration
u
Select the user you wish to use.
...
...
u
Tap the specified function on the function bar.
ð
The user is active.
ð
The user is marked up green.
SET
8.8
Installing the sample tube
NOTE
Removing is done in reverse sequence.
Precondition:
R
Make sure, that the end stop is in the correct
position for the sample tube in use. See
Chapter 8.10 "Preparing the end stop for the
sample tube size", page 73
u
Hold the sample tube with the tongs.