10 | Cleaning and servicing
Büchi Labortechnik AG
82/106
Operation Manual K-365 Dist Line
10.1
Regular maintenance work
Action
Daily
Weekly
Monthly
Yearly Additional information
Cleaning and servicing the
sample tube
1
Carry out this action before
every instrument use.
Cleaning the glass compo-
nents
1
Carry out this action after
every instrument use.
Cleaning and servicing the
hoses and hose connec-
tions
2
Cleaning and servicing the
sample tube seal
2
Replace the sample tube
seal twice per year
Cleaning and servicing the
splash protector
2
Replace the glass splash
protector after ~ 3500 dis-
tillations (twice per year).
Replace the plastic splash
protector once in 2 years.
Cleaning and servicing the
steam inlet and condenser
outlet tube
2
Replace the tube once per
year.
2
Cleaning and servicing the
warning and directive sym-
bols
2
Cleaning and servicing the
bridge splash protector to
condenser
2
Replace the bridge once
per year.
Cleaning and servicing the
dosing pumps
2
Replace the NaOH pump
once per year
Cleaning and servicing the
condenser
2
Cleaning and servicing the
steam generator
2
10.14.1 Calibrating the H₂O pump
2
10.14.2 Calibrating the acid pump
2
10.14.3 Calibrating the NaOH pump
2
2
2
1 - User; 2 - Operator
10.2
Cleaning and servicing the sample tube
u
Before using, check the sample tube for defects (cracks/splintering).
u
Prevent temperature shocks on the sample tubes.
ð
Temperature shocks may lead to breakage.