5 | Installation
Büchi Labortechnik AG
30/106
Operation Manual K-365 Dist Line
5.4
Installing the cooling water supply
5.4.1
Installing a tap water supply (option)
Precondition:
R
The tap water supply complies with the
specified parameters. See Chapter 3.5
"Technical data", page 24
R
Make sure that the instrument is not
connected to the power supply.
u
Install the inlet hose to the connection marked
Cooling Water / Chiller IN.
u
Install the drain hose to the connection marked
OUT Cooling Water / Chiller.
u
Put the other end of the hose in a sink.
5.4.2
Installing the recirculating chiller on the condenser (option)
NOTE
Place the hose in the rear cable duct.
Precondition:
R
The water supply complies with the tap water
specifications. See Chapter 3.5 "Technical
data", page 24
u
Make sure that the instrument is not connected
to the power supply.
u
Remove the connection at the condenser inlet.
u
Attach the cooling water hose from the chiller
to the condenser.