Büchi Labortechnik AG
Preparations for a determination | 8
Operation Manual K-365 Dist Line
69/106
8.6.18 Changing the dosing unit volume for a method (MultiDist T only)
NOTE
The availability depends on the selected titration type.
Specifications on the display:
u
Navigate to the specified submenu.
Methods
u
Select the method that you wish to edit.
...
u
Select the specified action.
...
Dosing Unit Vol.
u
Tap the specified function on the function bar.
EDIT
u
Enter the quantity for the volume.
...
u
Tap the specified function on the function bar.
ð
The volume is saved.
SAVE
8.6.19 Changing the titration stirring speed for a method (MultiDist T only)
NOTE
The availability depends on the selected titration type.
Specifications on the display:
Precondition:
R
A titration type is activated. See
u
Navigate to the specified submenu.
Methods
u
Select the method that you wish to edit.
...
u
Select the specified action.
...
Stirrer Speed Titr.
u
Tap the specified function on the function bar.
EDIT
u
Enter the stirring speed.
...
u
Tap the specified function on the function bar.
ð
The stirring speed is saved.
SAVE
8.6.20 Changing the titration start time for a method (MultiDist T only)
NOTE
The availability depends on the selected titration type.