6 | Description of the interface
Büchi Labortechnik AG
42/106
Operation Manual K-365 Dist Line
Action
Option
Explanation
[Current Time]
Time input
Enter in sequence: Minutes, hours.
Apply the settings by pressing
[Save]
.
Time zone
Choice of time zone
Select your time zone
[Keyboard Layout]
Choice of display key-
board on the interface
Select the keyboard layout
[Key Tones]
Off/On
Setting for audible signal in response
to input controls.
[Beep on Finish]
Off/On
Setting for audible signal by the end
of a determination.
[Beep on Error]
Off/On
Setting for audible signal by the end
of a determination.
[Display Brightness]
Enter setting
Display illumination level in %: 0 -
100
[Zero Amount Warn-
ing]
On/Off
The instrument gives a warning when
the entry for the weight is zero.
[Default Amount
Unit]
Select value
The following values are selectable:
g, mL
[MaxAccuracy
Mode]
On / Off
Compensation of the steam power
caused by voltage fluctuations.
[AutoDist Mode]
On/Off
Conditioning and distillation are per-
formed automatically.
[Level Detect. Max.
Time]
Enter value
The time after which the distillation
stops without level detection sensor
triggering.
[Pump Maintenance
Interval]
Select value
Selecting a frequency to carry out a
pump rinsing.
See Chapter 10.20 "Rinsing a pump",
page 90
[H₂O Dosage per
Keypress]
Enter value
Dosage volume which is applied by
pressing the related button.
[Acid Dosage per
Keypress]
Enter value
Dosage volume which is applied by
pressing the related button.
[NaOH Dosage per
Keypress]
Enter value
Dosage volume which is applied by
pressing the related button.
[Aspiration Time Re-
ceiver]
Select value
Aspiration time for the related but-
ton.
[Aspiration Time
Sample Tube]
Select value
Aspiration time for the related but-
ton.
[H₃BO₃ Dosage per
Keypress]
Enter value
Dosage volume which is applied by
pressing the related button.
[Steam Generator
Standby]
Enter value
Enter the time until the instrument
turns automatically into standby.