11 | Help with faults
Büchi Labortechnik AG
96/106
Operation Manual K-365 Dist Line
Problem
Possible cause
Action
Steam generator is
not reaching the
status Ready
Not enough water in the steam
generator.
u
Make sure, that the
installation H₂O supply for
steam generation is correct.
u
Make sure, that there is
enough water in the
canister.
u
Make sure, that the hose is
submerged in the water.
u
Check for leaks between
instrument and canister.
u
Check water level status in
the steam generator.
Display is black
The connection between in-
strument and display is inter-
rupted.
u
Check the connection cable
from the instrument to the
display.
u
No cooling water
flow
The cooling water supply is
blocked.
u
Make sure, that the hoses
are not bend.
u
Check for leaks between
instrument and cooling
water source.
Water flow sensor blocked.
u
Clean the hoses.
Dosing pump is not
feeding
Not enough liquid for feeding.
u
Make sure, that the
installation is correct.
u
Make sure, that there is
enough liquid in the
canister.
u
Make sure, that the hose is
submerged in the liquid.
u
Check for leaks between
instrument and canister.
u
Make sure that liquid is
inside the pump.
Aspiration is not
working
Leaks
u
Check the hoses connected
with the pump for leaks and
deterioration.
The aspiration pump is defect.
u
Check if the wheel inside the
pump is turning.
u