Büchi Labortechnik AG
Operation | 6
Operation Manual Inert Loop S-395
21/34
u
All following functions of the instrument are controlled by the spray drying
system.
6.3
Tasks during spray drying
u
Check the filling level of the solvent receiving vessel.
ð
If necessary empty it. See Chapter 7.2 "Empty the solvent receiving vessel",
page 22
u
Check the temperature.
6.4
Shutting down the instrument
Precondition:
R
The spray drying process is finished. See operation manual of the connected
spray dryer.
u
Switch off the spray gas. See operation manual of the connected spray dryer.
ð
The signal lamp oxygen is on.
u
Switch off temperature. See operation manual of the connected spray dryer.
u
Wait until the temperature is less than 70 °C.
u
Switch off the aspirator. See operation manual of the connected spray dryer.
ð
The signal lamp pressure is on.
u
Empty the solvent receiving vessel. See Empty the solvent receiving vessel
6.5
Switching off the instrument
Precondition:
R
The instrument is shut down. See Shutting down the instrument
u
Press the
Switch off
button. See Chapter 3.2.3 "Control elements", page 11.
u
Disconnect the communication cable.
u
Disconnect the power cable.
6.6
Setting condenser temperature
Increase temperature
u
Press the button
Set temperature UP
. See Chapter 3.2.3 "Control elements",
Decrease the temperature
u
Press the button
Set temperature DOWN