8 | Help with faults
Büchi Labortechnik AG
30/34
Operation Manual Inert Loop S-395
8
Help with faults
8.1
Error messages
NOTICE
Too frequent switching on the instrument
Too frequent switching on the instrument can cause an instrument damage.
u
Wait 10 minutes before restarting the instrument.
Fault code Fault
Remedy
E01
Temperature sensor
defective
u
Switch off the instrument.
u
Wait until the instrument is in ambient
temperature.
u
Remove dust and foreign objects from the
ventilation slots using compressed air or a
vacuum cleaner.
u
Switch on the instrument.
u
Contact BUCHI Customer Service.
E02
Temperature error
E04
Compressor pres-
sure fault
u
Switch off the instrument.
u
Wait until the instrument is in ambient
temperature.
u
Switch on the instrument.
u
Contact BUCHI Customer Service.
E05
Data error
u
Switch off the instrument.
u
Switch on the instrument.
u
Contact BUCHI Customer Service.
E06
Electronic circuitry
overheated
u
Switch off the instrument.
u
Wait until the instrument is in ambient
temperature.
u
Remove dust and foreign objects from the
ventilation slots using compressed air or a
vacuum cleaner.
u
Switch on the instrument.
u