Büchi Labortechnik AG
Cleaning and servicing | 7
Operation Manual Inert Loop S-395
23/34
u
Disconnect the solvent receiving vessel.
u
Remove the solvent receiving vessel from the
instrument.
u
Empty the solvent receiving vessel.
u
Install the solvent receiving vessel in revers
order.
7.3
Cleaning the housing
u
Wipe down the housing with a damp cloth.
u
If heavily soiled, use ethanol or a mild detergent.
u
Wipe down the display with a damp cloth.
7.4
Cleaning the ventilation slots
u
Remove dust and foreign objects from the ventilation slots using compressed air
or a vacuum cleaner.
7.5
Calibrating the oxygen analyzer
NOTE
The calibration should be performed with the sensor removed.
To avoid calibration errors, clean ambient air or certified 20.9% level oxygen should
be applied to the sensor.
u
If an ambient air calibration is performed, it is recommended that the oxygen
level is confirmed by a portable oxygen meter.