2 | Safety
Büchi Labortechnik AG
8/34
Operation Manual Inert Loop S-395
2.6.1
Faults during operation
If an instrument is damaged, sharp edges, glass splinters, moving parts or exposed
electrical wires can cause injuries.
u
Regularly check instruments for visible damage.
u
If faults occur, switch off the instrument immediately, unplug the power cord and
inform the operator.
u
Do not continue to use instruments that are damaged.
2.6.2
Glass breakage
WARNING
In the rare case of an abrupt massive leakage like a glass breakage a deflagration
can happen.
Risk of injury from flying glass splinters.
u
Leave the instrument immediately.
u
Stay in a safe distance for 1 minute.
u
Do not switch off or unplug the instrument.
u
Wait until the instrument changed independently into a safe state to get back.
2.7
Modifications
Unauthorized modifications can affect safety and lead to accidents.
u
Use only genuine BUCHI accessories, spare parts and consumables.
u
Carry out technical changes only with prior written approval from BUCHI.
u
Only allow changes to be made by BUCHI service technicians.
BUCHI accepts no liability for damage, faults and malfunctions resulting from
unauthorized modifications.
2.8
Oxygen sensor
The instrument is equipped with an oxygen measurement sensor to ensure a safe
operation when spraying combustible solvent. The inertization of the instrument is
continuously monitored. In case of an oxygen concentration above 6% the
instrument is able to immediately stop the solvent spraying. A safe state will be
reached with a maximum time delay of 15 to 25s. The actual time delay depends on
the drying gas flow rate.
The used sensor is SIL capable.