Büchi Labortechnik AG
Product description | 3
Operation Manual Inert Loop S-395
15/34
3.4.4
Installation site
The installation site meets the safety requirements. See Chapter 2 "Safety",
page 6
The installation site has a firm, level and nonslip surface.
The installation site has no obstacles (e.g. water taps, drains, etc.).
The installation site has an own mains outlet socket for the instrument.
The installation site is not exposed to external thermal loads, such as direct solar
radiation.
The installation site has enough space that cables / tubes can be routed safely.
The installation site meets the requirements for the connected devices. See
related documentation
The installation site meets the specifications according to the technical data (e.g.
weight, dimension, etc.). See Chapter 3.4 "Technical data", page 12
The installation site allows that the power supply can be disconnected at any time
in an emergency.
The installation site fits basic electromagnetic environment / Emission Class B.
Make sure there is free flow of air to the side of the instrument.
Do not place loose papers or cloths below or to the sides of the instrument, as
these could impede the air circulation if drawn in.
Put the instrument on the castors directly on the floor, do not use a pad.
Make sure the castor brakes are locked.
Place only BUCHI spray dryer on the instrument.
Do not place the instrument near vibration-sensitive devices.