BÜCHI Labortechnik AG
Installation | 5
Operation Manual Ultrasonic Package
15/34
5
Installation
5.1
Before installation
NOTICE
Instrument damage due to switching it on too early.
Switching on the instrument too early after transportation can cause damage.
u
Climatize the instrument after transportation.
5.2
Installation site
Place the ultrasonic controller on the product feed table of the Mini Spray Dryer
B-290.
NOTE
Make sure that the power supply can be disconnected at any time in an emergency.
5.3
Establishing electrical connections
NOTE
Observe the legal requirements when connecting the instrument to the power supply.
4
Use additional electrical safety features (e.g., residual-current circuit breakers) to
comply with local laws and regulations.
The power supply must fulfil the following conditions:
1. Provide the mains voltage and frequency specified on the type plate of the instru-
ment.
2. Be designed for the load imposed by the instruments connected.
3. Be equipped with suitable fuses and electrical safety features.
4. Be equipped with proper earthing.
NOTICE
Risk of property damage and diminished performance due to use of un-
suitable power cables.
The power supply cables supplied with the product by BUCHI precisely match the re-
quirements of the device. If other power cables that do not meet those requirements
are used, the device may be damaged and/or its performance diminished.
u
Use only the power supply cables supplied with the product or ordered separately
from BUCHI.
u
If using any other power supply cables, make sure that they match the specifica-
tions on the type plate.
u
Make sure that all connected devices are earthed.
u
If an extension lead is required, make sure that it is earthed and has a suitable
power rating.
u
Make sure that the power plug is freely accessible at all times.