8 | Help with faults
BÜCHI Labortechnik AG
30/34
Operation Manual Ultrasonic Package
8
Help with faults
8.1
No liquid delivery
u
Check the condition of the tube used in the peristaltic pump.
ð
Replace a worn out tube.
u
Adjust the peristaltic pump bed. See
"Operation Manual Mini Spray Dryer B-290 "
8.2
Liquid emerges unatomized
u
Check if the maximum flow rate is within the rated nozzle capacity.
u
Check if the feed viscosity and particle size of the slurry are within the specifica-
tions.
u
Turn off the peristaltic pump.
u
Switch the Ultrasonic Controller on/off for 5 times to get the liquid out of the ultra-
sonic nozzle, or use the damper by squeezing it.
u
Adjust the spray cone. See Chapter 6.2 "Finding the optimal spray cone", page 23
8.3
Error messages
Message
Explanation
Solution
Alarm
The nozzle is not con-
nected
u
Connect the nozzle
The nozzle temperature is
more than 100 °C
u
Increase the cooling
supply
LED lighting
The nozzle temperature is
more than 100 °C
u
Increase the cooling
supply