BÜCHI Labortechnik AG
Operation | 6
Operation Manual Ultrasonic Package
25/34
u
Wait 2 -3 minutes.
ð
The solvent flushes the remainings from the ultrasonic nozzle.
u
Tap the button main switch peristaltic pump.
u
Switch the On/Off master switch of the ultrasonic controller to Off.
u
Tap the button main switch heater.
u
Wait until the glassware cooled down.
u
Tap the button main switch aspirator.
u
Set the air flow on the rotameter to 0.
u
Remove the product from the product collection vessel.
6.3.6
Shutting down the instrument
Precondition:
R
The spray drying process is finished. See Chapter 6.3.5 "Ending a spray drying
process in open mode", page 24
u
Switch the On/Off master switch of the Mini Spray Dry B-290 to Off.
u
Clean the ultrasonic nozzle. See Cleaning and servicing
u
Clean the Mini Spray Dryer B-290. See
"Operation Manual Mini Spray Dryer B-290
"
6.4
Carrying out a spray drying precess in closed loop mode
There are two different modes closed mode:
Closed mode with B-295
and B-296
Closed mode with B-295
Instruments required
Mini Spray Dryer B-290
Mini Spray Dryer B-290
Inert Loop B-295
Inert Loop B-295
Dehumidifier B-296
6.4.1
Tasks before starting spray drying in closed mode
Precondition:
R
u
Prepare 50 mL pure solvent.
u
Prepare the sample.
u
Place sample and solvent on top of the ultrasonic controller.
6.4.2
Preparing the instrument for closed mode
Low drying temperature
ca. 80 °C
High drying temperature
ca. 220 °C
Time required:
approx. 15 min
approx. 30 min
u
Switch the On/Off master switch of the Mini Spray Dryer B-290 to On.
ð
The instrument is starting up.
ð
The Inert Loop B-295 is starting up.
u
Select the condenser temperature at the Inert Loop B-295.
u
For closed mode with B-295 and B-296 switch the On/Off master switch of the De-
humidifier B-296 to On.
u
Set the flow on the rotameter between 30 - 40 mm.
ð
The cooling gas cools the ultrasonic nozzle.
u
Select the aspirator power.