5 | Installation
BÜCHI Labortechnik AG
18/34
Operation Manual Ultrasonic Package
u
Attach the assembled feeding tube to the noz-
zle.
u
Attach the inlet tube in place with the union
nut.
u
Install the feed tube at the peristaltic pump.
See operation manual Mini Spray Dryer
B-290.
u
Adjust the peristaltic pump bed. See
"Opera-
tion Manual Mini Spray Dryer B-290 "
5.6
Installing the ultrasonic nozzle in closed mode
DANGER
Incorrect cooling gas in closed mode e.g. compressed air
Incorrect cooling gas can lead to an potential explosive environment.
u
Use inert gas.
NOTICE
Water as cooling medium
Water as cooling medium can cause an instrument damage.
u
Use compressed air.
u
Insert the ultrasonic nozzle into the heater ele-
ment of the Mini Spray Dryer B-290.
u
Attach the ultrasonic nozzle with the ultrasonic
nozzle cable on the ultrasonic controller.